Calibration Factor Calculator
Calculate a proposed calibration factor from the current factor, target value and measured value.
Input Values
Results
Apply a revised factor only under the laboratory’s approved calibration procedure.
Formula And Worked Example
Formula
New factor = current factor × target ÷ measured.
Worked Example
A factor of 250, a target of 100 and a measured value of 80 give 250 × 100 ÷ 80 = 312.5. Under the proportional model, this represents an increase by a multiplier of 1.25.
Inputs And Units
Current factor; target concentration or value; measured concentration or value. The target and measured values must be in the same unit; all three inputs must be positive.
Interpretation And Limits
The result is a proposed mathematical factor, not authorization to recalibrate an analyzer. An out-of-control result may arise from reagent deterioration, preparation error, pipetting, temperature, instrument malfunction or other causes. Investigate these before any adjustment. Apply changes only when the method permits them, using authorized calibrators, documented procedures and verification before patient reporting. A factor cannot repair nonlinearity, an offset or poor precision.