Precision and Bias Verification: A Practical Introduction
A practical introduction to precision, CV%, bias, method verification design, acceptance criteria, and documented decisions.

Before a laboratory begins routine reporting on a new or changed quantitative test system, it needs evidence that the method performs acceptably under local conditions. Two important questions are: 'How repeatable are the results?' and 'How far might the results differ from an appropriate reference or comparison method?' These relate to precision and bias.
Verification is not identical to a manufacturer's full method validation. The study design and acceptance criteria depend on the method, intended use, whether it is a modified or unmodified procedure, and applicable requirements. Plan the study before collecting results, and get appropriate technical approval.
Precision: how closely results agree
Precision describes variation among repeated measurements under specified conditions. A set of results can be tightly grouped yet consistently too high; that would be precise but potentially biased. Common measures are standard deviation (SD) and coefficient of variation (CV%):
CV% = (SD / mean) × 100
For example, if repeated results have a mean of 100 units and an SD of 2 units, CV is 2%. At means near zero, CV% can become misleading or undefined; evaluate variation using an appropriate alternative measure for that concentration range. This is only an arithmetic illustration, not an acceptance limit. A 2% CV may be excellent for one analyte and unacceptable for another depending on the clinical requirement.
Design the experiment around clinical decisions
Use material representing important concentration levels, including decision points where practical. Define the number of runs, replicates, days, operators, and instruments in the study plan. Some studies measure short-term repeatability; others capture within-laboratory variation over different days and runs. Do not claim to have verified long-term precision from a few measurements made in one hour.
Record sample handling, lot information, environmental or maintenance changes, and exclusions. Calculate statistics using the prespecified method. If an outlier appears, investigate it rather than deleting it because the CV would otherwise look better.
Bias: how results differ from a suitable target
Bias is a systematic difference from an appropriate reference value or comparative method. It is often expressed in units or percent, depending on the measurand. A simple percentage difference is:
Percent difference = (test method − comparison value) / comparison value × 100
This simple calculation is undefined when the comparison value is zero and can be misleading at very low concentrations. It does not by itself establish method bias across the reportable range. Method-comparison studies should use suitable statistical analyses and cover the clinically relevant measurement range; one specimen is not enough.
Select reference materials and methods carefully. Proficiency-testing peer-group averages may be useful for monitoring, but they are not automatically equivalent to a traceable reference value. Material commutability and method grouping matter when evaluating systematic differences.

Set acceptance criteria before reviewing data
Criteria may be based on clinical risk, relevant quality specifications, regulatory requirements, method performance claims, and the intended use of the assay. Avoid inventing a universal acceptable CV or bias percentage. A method can meet a convenient numeric target and still perform poorly near a critical clinical decision threshold.
Present the data in a clear summary: material or specimen level, mean, SD/CV, comparison method or target, observed difference, acceptance criterion, and pass/fail decision. The calculation sheet should be linked to raw observations and signed off by authorized staff.
Investigate failures instead of averaging them away
If imprecision is excessive, review pipetting, reagent preparation, calibration, instrument maintenance, stability, and environmental conditions. If bias appears concentration-dependent, consider the measurement range, calibrator traceability, matrix effects, and method-specific limitations. Correct the cause and repeat the appropriate part of the verification under a documented plan.
Keep a record of what changed and when re-verification is needed. Major repairs, new instrument platforms, significant method modifications, or other changes can trigger additional evaluation under local policies; not every routine software update automatically needs the same entire verification study.
A practical manager's checklist
Before approving the method, ask whether the study covers relevant concentrations, whether the number of measurements is justified, whether limits were fixed in advance, whether raw data are retained, and whether the final decision is understandable to someone who did not perform the calculations.
The purpose is confidence that patient results are fit for clinical use—not producing a neat spreadsheet that always ends in 'pass.'
Example: precise results that may still be wrong
A laboratory tests a control material repeatedly and obtains values clustered very closely together. The calculated CV is low, so repeatability looks good. But the mean may still sit above an appropriate reference value or established comparison method. In that case, precise measurements could be consistently biased. Reviewing only the CV would miss the second problem.
A useful verification report includes a table for each relevant concentration: the sample or control level, measurement count, observed mean, SD, CV, comparison value where applicable, observed difference, acceptance criterion, and final decision. The plan should specify how low-level results and concentration-dependent differences are interpreted. A simple percent difference becomes unstable when the comparison value approaches zero; choosing the right comparison method matters.
Avoid: selecting acceptance limits after seeing results, removing inconvenient observations without justification, describing a peer-group mean as an unquestionable reference truth, or declaring an entire measurement range verified using one easy sample. The aim is clinical fitness for use, not a neat pass label.
Common questions
Is low CV proof that a method is accurate?
No. Low CV indicates good precision under the conditions studied, but a method can give closely grouped results that are systematically too high or low. Assess bias or agreement separately.
Can one control level verify the whole reportable range?
Usually not. A well-designed study considers clinically important concentrations and the method range. The exact levels and replicates depend on the verification protocol and applicable standards.
Reference leads
CLSI EP15 (user verification of precision and bias), EP05 (precision), EP09 (method comparison); ISO 15189:2022.